Use the serial line to select and scale the
analog output quantities. Connect HMP155 to the PC and open the terminal connection between
your computer and the sensor.
| HMP155 must be in the adjustment mode for
the AMODE command to work. To switch the probe to the
adjustment mode, hold down the ADJ button until a green LED lights
up. |
Check the analog output modes with the AMODE command.
Table 1. Analog output modes
Analog output mode |
Description |
3 |
0 … 1 V |
4 |
0 …5 V |
5 |
0 … 10 V |
| Analog output calibration is
valid only for the factory settings. If you change the analog output mode with AMODE command, you must continue by entering the ACAL command, see ACAL command. |
Example
>amode 3 3
Ch0 : 0 ... 1V
Ch1 : 0 ... 1V
>