AMODE command - HMP155

HMP155 User Guide

Document code
M210912EN
Revision
G.1
Language
English
Product
HMP155
Document type
User guide

Use the serial line to select and scale the analog output quantities. Connect HMP155 to the PC and open the terminal connection between your computer and the sensor.

HMP155 must be in the adjustment mode for the AMODE command to work. To switch the probe to the adjustment mode, hold down the ADJ button until a green LED lights up.

Check the analog output modes with the AMODE command.

Table 1. Analog output modes
Analog output mode Description
3 0 … 1 V
4 0 …5 V
5 0 … 10 V
Analog output calibration is valid only for the factory settings. If you change the analog output mode with AMODE command, you must continue by entering the ACAL command, see ACAL command.

Example

>amode 3 3
Ch0 : 0 ... 1V
Ch1 : 0 ... 1V
>