Push-button calibration - HMP155

HMP155 User Guide

Document code
M210912EN
Revision
G.1
Language
English
Product
HMP155
Document type
User guide
Push-button calibration is not available for the additional temperature probe.

You can calibrate and adjust HMP155 with the adjustment buttons on the probe. Carry out the two-point humidity adjustment by using 2 relative humidity references: for example, saturated salt points 11 %RH (LiCl) and 75 %RH (NaCl). The difference between the humidity reference points must be at least 30 %RH. The difference between the temperature reference points must be at least 30 °C.

In push-button calibration, the probe outputs are by default relative humidity in the range 0 … 100  %RH in CH1 and temperature in the range -40 … +60 °C (-40 … +140 °F) in CH2.

Before attempting calibration or adjustment, set the probe to the adjustment mode by pressing and holding the ADJ button until a green LED lights up. This turns off the heating and the additional temperature probe if you are using either of them. After that HMP155 is ready to be checked and calibrated and adjusted. Because the additional temperature probe is turned off during the adjustment mode, you do not need to insert it into the same reference conditions with the humidity probe.