HMP155 has an 8-pin M12 male connector on the bottom of the probe.
Pin | Open-end wire color | Passive output | Active output |
---|---|---|---|
1 | White | Pt100 | VOUT1 |
2 | Brown | RHOUT 0 … 1 V / RS-485-B | RS-485-B |
3 | Green | Pt100 | AGND |
4 | Yellow | Pt100 | VOUT2 |
5 | Gray | Pt100 | - |
6 | Pink | AGND / RS-485-A | RS-485-A |
7 | Blue | VCC | VCC |
8 | Red | GND | GND |
- | Black | Shield | Shield |
The shield is not a pin but a conductor in the connection cable.
The left side shows the HMP155D block diagram. The right side shows an example of connections.
Make sure that the analog ground (AGND) and the power ground (GND) are isolated from each other. If they are not isolated, analog T and RH readings may be disturbed. |
The left side of the figure shows the HMP155A/E block diagram. The right side shows an example of connections.
Make sure that the analog ground (AGND) and the power ground (GND) are isolated from each other. If they are not isolated, analog T and RH readings may be disturbed. |