Mounting HMP155 inside radiation shield DTR503A - HMP155

HMP155 User Guide

Document code
M210912EN
Revision
G.1
Language
English
Product
HMP155
Document type
User guide
  • 3‑mm, 4‑mm, and 5‑mm Allen keys (provided)
  • 10‑mm wrench
Figure 1. HMP155 and DTR503A installation accessories
1
Radiation shield (DTR503A)
2
Mounting plate
3
Support plate
4
Pole or support Ø 30 mm (1.18 in) or 60 mm (2.36 in)
5
U‑bolt
6
Hex screw M4×16 DIN912 (2 pcs). Tightening torque 5 Nm.
7
Spring washer B4 DIN127 (2 pcs)
8
Flat washer A4,3 DIN125 (2 pcs)
9
HMP155
10
Hex screw M5×12 DIN912 (4 pcs)
11
Spring washer B5 DIN127 (4 pcs)
12
Flat washer A5,3 DIN125 (4 pcs)
13
Nut M6 DIN934 (2 pcs)
14
Spring washer B6 DIN127 (2 pcs)
15
Flat washer A6,4 DIN125 (2 pcs)
  1. Attach the mounting plate to the support plate.
  2. Secure the mounting plate to the pole.
  3. Attach the radiation shield to the mounting plate.
  4. Remove the protective cover from the probe.
  5. Lead the sensor cable through the connector cover, and attach the cable to the probe.
  6. Loosen the compression fitting with your fingers and insert the probe inside the ring.
    Pay attention to the rubber insert.
    1
    Compression fitting
    2
    Rubber insert
  7. Tighten the compression fitting until it locks in place.
    CAUTION Avoid using excessive force.
  8. Push the connector cover over the probe connector.
  9. Lead the sensor cable along the support arm and attach the cable to the support arm with cable ties. Connect the sensor cable to the M12 connector in HMP155.