Use the FILT command to set the filtering level. The default setting is 0.8. The output is calculated based on the following formula: [( new result * filt) + ( old result * (1.0 - filt))]. With FILT value 1, the probe only takes the latest measurement into account. With FILT value 0.1, a new output is a combination of the previous output (90 %) and the latest measurement (10 %).
FILT xxx
Parameter | Description |
---|---|
xxx | 0.1 … 1 1.0 = no filter 0.1 = about 16 moving average |
Example
>filt
Filter : 0.800 ?
>