Use the CT command for one-point or two-point serial line temperature calibration and adjustment of the T sensor. The CT command changes the user adjustment parameters of temperature sensor T.

You can use the CT command only in the adjustment mode.

Example of two-point adjustment:

>ct
T : 18.6038 1. ref ?
T : 18.6068 1. ref ?
T : 18.6098 1. ref ? 19.0
Press any key when ready ...
T : 49.5176 2. ref ?
T : 49.5176 2. ref ?
T : 49.5176 2. ref ? 50.0
OK
>

OK indicates that the adjustment has been successful.