You can enable or disable the extra sensor heating with the XHEAT command.

>xheat on
Extra heat     : ON
>xheat off
Extra heat     : OFF
>

To configure extra sensor heating, use the XHEAT command without parameters. Type the values after the question mark. The following table lists the available ranges.

Available ranges Description
0 ... 100 %RH (default: 95 %RH) Extra heating RH limit (heating function starts up above the setpoint)
0 ... 200 °C (default: 100 ºC) Extra heating temperature
0 ... 255 s (default: 30 s) Extra heating time

To see the current values, use the XHEAT * command.

After the defined XHEAT duration, there is an additional cooling time of 10 seconds before outputs are updated.

Example


>xheat *
Extra heat     :      OFF
RH limit       :       95 %RH
Temperature    :      100 'C
Duration       :       30  s
>