You can enable or disable the extra sensor heating with the XHEAT command.
>xheat on
Extra heat : ON
>xheat off
Extra heat : OFF
>
To configure extra sensor heating, use the XHEAT command without parameters. Type the values after the question mark. The following table lists the available ranges.
Available ranges | Description |
---|---|
0 ... 100 %RH (default: 95 %RH) | Extra heating RH limit (heating function starts up above the setpoint) |
0 ... 200 °C (default: 100 ºC) | Extra heating temperature |
0 ... 255 s (default: 30 s) | Extra heating time |
To see the current values, use the XHEAT * command.
After the defined XHEAT duration, there is an additional cooling time of 10 seconds before outputs are updated. |
Example
>xheat *
Extra heat : OFF
RH limit : 95 %RH
Temperature : 100 'C
Duration : 30 s
>