Extra sensor heating (Xheat) option - HMP155

HMP155 User Guide

Document code
M210912EN
Revision
G.1
Language
English
Product
HMP155
Document type
User guide

You can define the humidity sensor heating temperature and the duration of the heating. Sensor heating starts when the relative humidity of the measuring environment reaches the RH value (RH limit) you have set.

After the heating cycle, a new sensor heating cycle is performed if the predefined conditions are reached again.

During the sensor heating, the outputs are locked to the values measured before the heating cycle.