HMP155 specifications - HMP155

HMP155 User Guide

Document code
M210912EN
Revision
G.1
Language
English
Product
HMP155
Document type
User guide
Table 1. Humidity measurement performance
Property Description/value
Sensor

HUMICAP®R2, 180R and INTERCAP for typical applications

HUMICAP®R2C, 180RC and INTERCAPC for applications with chemical purge and/or warmed probe

Observation range 0 … 100 %RH
Response time at +20 °C (+68 °F) in still air with sintered Teflon filter

63 %: 20 s

90 %: 60 s

Factory calibration uncertainty at +20 °C (+68 °F)1

±0.6 %RH (0 … 40 %RH)

±1.0 %RH (40 … 95 %RH)

Accuracy (including non-linearity, hysteresis, and repeatability)
At +15 … +25 °C (+59 … +77 °F)

±1 %RH (0 … 90 %RH)

±1.7 %RH (90 … 100 %RH)

At −20 … +40 °C (−4 … +104 °F) ±(1.0 + 0.008 × reading) %RH
At −40 … −20 °C (−40 … −4 °F) ±(1.2 + 0.012 × reading) %RH
At +40 … +60 °C (+104 … +140 °F) ±(1.2 + 0.012 × reading) %RH
At −60 … −40 °C (−76 … −40 °F) ±(1.4 + 0.032 × reading) %RH
Table 2. Temperature measurement performance
Description Value
Sensor Pt100 RTD element, Class F 0.1 IEC 60751
Observation range −80 … +60 °C (−112 … +140 °F)
Response time for additional temperature probe in 3 m/s (7 mph) air flow

63 %: < 20 s

90 %: < 35 s

Accuracy with voltage output
At −80 … +20 °C (−112 … +68 °F) ±(0.226 - 0.0028 × temperature) °C
At +20 … +60 °C (+68 … +140 °F) ±(0.055 + 0.0057 × temperature) °C
Accuracy with passive (resistive) output
According to Tolerance Class AA IEC 607512 ±(0.1 + 0.0017 × |temperature|) °C
Accuracy with RS‑485 output
At −80 … +20 °C (−112 … +68 °F) ±(0.176 - 0.0028 × temperature) °C
At +20 … +60 °C (+68 … +140 °F) ±(0.07 + 0.0025 × temperature) °C
Figure 1. HMP155 accuracy over temperature range: voltage and RS-485
Table 3. Dew point calculation specifications
Property Description/value
Accuracy at −20 … +40 °C (−4 … +104 °F) ±0.6 at 90 … 100 %RH
Table 4. Quantities measured by HMP155
Quantity Abbreviation Metric unit Imperial unit
Relative humidity RH %RH %RH
Temperature T °C °F
Additional T‑probe temperature Ta °C °F
The calculated quantities are not available in the passive output version.
Table 5. Quantities calculated from measured quantities
Quantity Abbreviation Metric unit Imperial unit
Dew point / Frost point temperature (Td/f) TDF °C °F
Dew point temperature (Td) TD °C °F
Mixing ratio (x) X g/kg gr/lb
Wet bulb temperature (Tw) TW °C °F
Table 6. Operating environment
Property Description/value
Operating temperature for humidity measurement

−80 … +60 °C (−112 … +140 °F)

Storage temperature

−80 … +60 °C (−112 … +140 °F)

Operating humidity

0 … 100 %RH

IP rating

IP66: Dust-tight. Protected from powerful water jets from any direction.

Table 7. Inputs and outputs
Property Description/value
Operating voltage 7 … 28 V DC
Minimum operating voltage

0 … 1 V output or RS‑485: 7 V

0 … 5 V output, or warmed probe: 12 V

0 … 10 V output, chemical purge, or XHEAT: 16 V

Outputs

Voltage output: 0 … 1 V, 0 … 5 V, 0 … 10 V

Resistive Pt100 4‑wire connection

RS‑485

Average power consumption (+15 V DC, load 100 kΩ)

0 … 1 V output: < 3 mA

0 … 10 V output: +0.5 mA

RS-485: < 4 mA

During chemical purge: Maximum 110 mA

With warmed probe: Maximum 150 mA

Settling time at startup

Voltage output: 2 s

RS-485: 3 s

Table 8. Mechanical specifications
Property Description/value
Dimensions (H × W) 279 × 40 mm (10.9 × 1.6 in)
Weight 93 g (3.25 oz)
Materials
Filter

Sintered Teflon or membrane

Housing Polycarbonate (PC)
Additional temperature probe Stainless steel AISI 316L
Table 9. Compliance
Property Description/value

EU directives and regulations

EMC Directive (2014/30/EU)

RoHS Directive (2011/65/EU) amended by 2015/863

EMC compatibility

EN 61326‑1, Electrical equipment for measurement, control and laboratory use - EMC requirements; industrial environment

CISPR 32 / EN 55032, Class B

RoHS

EN IEC 63000:2018

1 Defined as ±2 standard deviation limits. Small variations possible (see also the calibration certificate).
2 Tolerance Class AA IEC 60751 corresponds to IEC 751 1/3 Class B