Most of the following commands are relevant only in the digital output version of HMP155. To issue a command, type it on your computer and press Enter. |
Command | Default setting | Description |
---|---|---|
R | - | Start continuous output |
S | - | Stop continuous output |
INTV [0 ... 255 S/MIN/H] | 2 s | Set continuous output interval (RUN mode) |
SEND [0 ... 99] | - | Output reading once |
SMODE [STOP/RUN/POLL/SEND] | STOP | Set serial interface mode |
OPEN [0 ... 99] | - | Open temporary connection to POLL mode device |
CLOSE | - | Close temporary connection (back to POLL mode) |
SDELAY [0 … 255] | - | View or set user port (RS-485) answer minimum delay |
SERI [baud p d s] | - | User port settings (default: 4800 E 7 1) baud: 300 ... 115200 |
ADDR [0 ... 99] | 0 | Set probe address (POLL mode) |
Command | Description |
---|---|
FORM | Set the output format of SEND and R commands |
LASKE 6 | Sets the default output format used in Vaisala weather stations AWS310, AWS810, and RWS200 (non-heated probe option) |
LASKE 7 | Sets the default output format used in Vaisala weather stations AWS310, AWS810, and RWS200 and Vaisala Forward Scatter Sensor FD70 (heated probe option) |
TIME [HH MM SS] | Set or display time |
UNIT | Select metric or non-metric output units |
Command | Description |
---|---|
PUR | Set automatic chemical purge |
Command | Description |
---|---|
# | Force connection settings 19200 baud, N, 8, 1 |
ACAL [0/1] | Analog output calibration and adjustment |
CDATE | Set calibration date |
CRH | One or two-point relative humidity calibration and adjustment. |
CT | One or two-point temperature calibration and adjustment |
CTA | One or two-point additional temperature probe calibration and adjustment. |
FCRH | One or two-point relative humidity calibration and adjustment after sensor change |
CTEXT | Type text to calibration information field |
L | Display user adjustment parameters |
LI | Display new values for user adjustment parameters |
Command | Description |
---|---|
AMODE | View analog output modes |
ASEL | Set or view analog output quantities and scaling |
AERR | Change analog error output values |
ATEST | Force analog outputs to entered values |
Command | Description |
---|---|
? | Output information about the device |
?? | Output information about the device in POLL mode |
ERRS | List present probe errors |
FILT [0.1 … 1] | Set the result filtering |
HELP | List available commands |
PRES [hPa] | Set the value for pressure compensations |
RESET | Reset the probe |
VERS | Display software version information |
XHEAT | Sensor heating |
XPRES [bar] | Set temporary value for pressure compensations |