Cleaning window - CL61

CL61 User Guide

Document code
M212475EN
Revision
E
Language
English
Product
CL61
Document type
User guide
  • Soft, lint-free cloth
  • Suitable glass cleaning substance, such as isopropyl alcohol mixed with water
WARNING Viewing the laser output with telescopic optical instruments (for example, telescopes or binoculars) may pose an eye hazard. Do not direct the beam into an area where such instruments are likely to be used.
WARNING Only qualified maintenance personnel may perform maintenance procedures. Make sure unauthorized persons cannot access the work area during service operations.

The status message and netCDF files include a warning when the window is contaminated.

If the blower cannot remove the contamination, the ceilometer issues a warning indicating that you must clean the window manually.

  1. Make sure that the measurement unit door is closed.
    1
    Window
    2
    Measurement unit door
  2. Rinse the window with clean water to remove coarse grains.
    CAUTION Do not use pressure cleaners.
    CAUTION Avoid spraying water into the window blower.
  3. Clean the window with a soft, lint-free cloth moistened with a mild detergent.