Make sure that you have the replacement spare part available.
Replacing the window includes adjusting the
clean window reference. Set the new clean window reference for a clean and dry window, in
clear weather, and ideally with the termination hood placed on top of the device. The clean
window calibration ensures that the window condition diagnostics respond appropriately to
dirt accumulation.
Switch off AC (mains) power.
WARNING To make sure that
there is no AC (mains) power available for the system, disconnect the AC (mains) cable
(AC in).
1
AC (mains) power switch
2
AC (mains) power cable connector AC in
Clean the area around the window.
Remove the screws from around the window, and remove the window and the gasket.
1
Screws (12 pcs)
2
Slot for LED
3
Gasket, flat side up
Attach the gasket with the flat side up.
Attach the replacement window. Make sure to tighten all the screws.
Avoid touching the window with your bare
hands as it may get contaminated or scratched.
Make sure to place the window so that
the slot for the LED is on the bottom right corner.
Connect the power.
Change the user level.
level2
When the device has been running for 5 minutes, set a new clean window reference.
modeadjustcleanwindow
Type Y to
continue.
Setting the new reference takes a couple
of minutes. Wait for the message that this was successful.