Configuring location information - CL61

CL61 User Guide

Document code
M212475EN
Revision
E
Language
English
Product
CL61
Document type
User guide
  • Laptop
  • Maintenance cable

When starting to use CL61, define the device identifier and device location information. This information helps to distinguish several devices from each other and provides metadata for the measurements.

The following table lists the configurable unit metadata settings. Use the settings that are suitable for you.

Table 1. Unit metadata
Parameter Description Value / Range get set
cl.unit.alias1 Alias name 1 Freeform text 0 1
cl.unit.latitude

Latitude

Unit: degrees

−90.000000 … 90.000000 0 1
cl.unit.longitude

Longitude

Unit: degrees

−180.000000 ... 180.000000 0 1
cl.unit.altitude

Altitude

Unit: meters

Resolution: full meters

Use negative values (−) for values below sea level.

−999 ... 9999 0 1
cl.unit.azimuthangle

Measurement azimuth angle

Reference direction: north

Unit: degrees

0 ... 360 0 1
  1. Establish a maintenance connection to the device.
  2. Change the user level.
    level 1
  3. Configure the device identifier.
    set cl.unit.alias1 <alias>
  4. To set the device location, type the latitude, longitude, and altitude information.
    set cl.unit.latitude <latitude>
    set cl.unit.longitude <longitude>
    • For latitude the range is −90.0000 ... 90.0000.

    • For longitude the range is −180.0000 ... 180.0000.

    set cl.unit.altitude <altitude>

    For altitude, the resolution is full meters. Use negative value (−) for values below sea level.

  5. Set the azimuth angle.
    set cl.unit.azimuthangle <azimuth angle>