The following tables list the CL61 netCDF dimensions and variables. For the full list of attributes, types, and dimensions associated with each variable, see NetCDF example file.

Table 1. NetCDF dimensions
Dimension Description
time Temporal grid of the measurements, length unlimited.
range Range grid vector with a resolution of 4.8 meters and a length of 3276.
layer Vector that gives the number of the observed cloud layer, ordered from lowest to highest with length 5.
Table 2. Variables in netCDF file
Variable Attribute Value/Description

cloud_base_heights

cloud_base_heights:units

m, ft

cloud_base_heights:long_name

Heights (range) of the detected cloud bases

cloud_base_heights:coordinates Time layer longitude latitude

vertical_visibility

vertical_visibility:units

m, ft

vertical_visibility:long_name

Visibility in the direction of the instrument beam

vertical_visibility:coordinates Time longitude latitude

p_pol

p_pol:units

1/(m*sr)

p_pol:long_name

Parallel-polarized component of the backscattered light

p_pol:coordinates Time range longitude latitude
p_pol:averaging_time_in_seconds 60

x_pol

x_pol:units

1/(m*sr)

x_pol:long_name

Cross-polarized component of the backscattered light

x_pol:coordinates Time range longitude latitude
x_pol:averaging_time_in_seconds 60

beta_att

beta_att:units

1/(m*sr)

beta_att:long_name

Attenuated volume backscatter coefficient

beta_att:coordinates Time range longitude latitude
beta_att:averaging_time_in_seconds 60

linear_depol_ratio

linear_depol_ratio:long_name

Linear depolarization ratio of the backscatter volume

linear_depol_ratio:coordinates Time range longitude latitude
linear_depol_ratio:averaging_time_in_seconds 60

time

time:units

Seconds since 1970-01-01 00:00:00.000

time:long_name Time
time:axis T
time:standard_name time
time:cf_role Profile_id
time:comment

End of the averaging period

range

range:units

m, ft

range:long_name

Measurement distance from the instrument in the direction of the transmitted laser beam

range:axis Z
range:positive Up

layer

layer:units

Layer

layer:long_name

Number of the observed cloud layer (1 ... 5)

longitude

longitude:units

Degrees east

longitude:long_name

Longitude

longitude:standard_name longitude

latitude

latitude:units

Degrees north

latitude:long_name

Latitude

latitude:standard_name latitude

elevation

elevation:units

m

Always in meters

elevation:long_name Elevation
elevation:standard_name ground_level_altitude
elevation:comment

Measurement site height above or below a fixed reference point, most commonly a reference geoid

azimuth_angle

azimuth_angle:units

Degrees

azimuth_angle:long_name

Measurement azimuth angle

azimuth_angle:standard_name sensor_azimuth_angle
azimuth_angle:comment

Reference direction: north

beta_att_sum

beta_att_sum:units

1/(10^4*sr)

beta_att_sum:long_name

Scaled integral of the attenuated volume backscatter coefficient

beta_att_noise_level

beta_att_noise_level:long_name

Unitless number describing the noise level of the attenuated volume backscatter coefficient

tilt_correction

tilt_correction:long_name

Tilt correction

tilt_correction:comment
  • 1 = on
  • 0 = off

tilt_angle

tilt_angle:units

Degrees

tilt_angle:long_name

Instrument tilt angle from the vertical

tilt_angle:standard_name zenith_angle

height_offset

height_offset:units

m

Always in meters

height_offset:long_name

Instrument height offset to reference level.

height_offset:comment

Positive, if the instrument is placed, for example, on the roof of a building. Negative, if the instrument is placed below the ground level altitude, for example, in a pit. This value is added to the cloud base height results.

airplane_filter_max_range

airplane_filter_max_range:units

m, ft

airplane_filter_max_range:long_name

Airplane filter max. range.

airplane_filter_max_range:comment

User configured value, 0 for not in use, otherwise the configured range.

sky_condition_total_cloud

_cover

sky_condition_total_cloud_cover:units

Oktas

sky_condition_total_cloud_cover:long_name

Total amount of cloud cover.

sky_condition_total_cloud_cover:comment

Aggregated across layers.

sky_condition_total_cloud_cover:coordinates Time longitude latitude

sky_condition_cloud_layer

_covers

sky_condition_cloud_layer_covers:units

Oktas

sky_condition_cloud_layer_covers:long_name

Amount of cloud cover in different cloud layers.

sky_condition_cloud_layer_covers:coordinates Time layer longitude latitude
sky_condition_cloud_layer_covers:comment

For up to 5 layers

sky_condition_cloud_layer

_heights

sky_condition_cloud_layer_heights:units

m, ft

sky_condition_cloud_layer_heights:long_name

Height of different cloud layers.

sky_condition_cloud_layer_heights:coordinates Time layer longitude latitude
sky_condition_cloud_layer_heights:comment

For up to 5 layers

cloud_penetration_depth

cloud_penetration_depth:units

m, ft

cloud_penetration_depth:long_name

Cloud penetration depth in the direction of the instrument beam

cloud_penetration_depth:coordinates Time layer longitude latitude

cloud_thickness

cloud_thickness:units

m, ft

cloud_thickness:long_name

Cloud thickness in the direction of the instrument beam

cloud_thickness:coordinates Time layer longitude latitude

precipitation_detection

precipitation_detection:long_name

Detection of ground reaching precipitation

precipitation_detection:comment
  • 1 = detected
  • 0 = not detected
precipitation_detection:coordinates Time longitude latitude

fog_detection

fog_detection:long_name

Detection of fog

fog_detection:comment
  • 1 = detected
  • 0 = not detected
fog_detection:coordinates Time longitude latitude

receiver_gain

receiver_gain:long_name

Receiver gain status

receiver_gain:comment
  • 1 = high-gain
  • 0 = low-gain

range_resolution

range_resolution:units

m, ft

range_resolution:long_name

Range resolution.

range_resolution:comment

Distance between consecutive profile elements.

cloud_calibration_factor

cloud_calibration_factor:long_name

Factory cloud calibration value.

cloud_calibration_factor:comment

Instrument-specific beta_att calibration value measured at the factory.

cloud_calibration_factor_user

cloud_calibration_factor_user:long_name

User set cloud calibration value.

cloud_calibration_factor_user:comment

Instrument-specific beta_att calibration value measured by the user, same as the factory value by default.

overlap_function

overlap_function:long_name

Instrument-specific overlap function.

overlap_function:comment

Shares the vertical resolution of profiles.

Table 3. Variables, group monitoring
Variable Attribute Description
time time:units

Seconds since 1970-01-01 00:00:00.000

The timestamp of the monitoring values. Common reporting frequency with profiles.

time:long_name Time
time:axis T
time:standard_name time
window_condition window_condition:units

Percent

Window condition percentage between 100 and 0 representing a clean and a completely dirty window, respectively

window_condition:long_name Window condition
window_condition:comment 100 for a clean, 0 for a totally dirty window
laser_power_percent laser_power_percent:units

Percent

Point measurement of the laser output power, with a PIN photodiode at CLL board

laser_power_percent:long_name Laser power percent
background_radiance background_radiance:long_name

Background radiance

Measurement of the amount of sky background radiance, at the receiver field-of-view, arbitrary units

background_radiance:range [0 1747]
internal_temperature internal_temperature:units Celsius

Optics unit - circuit board temperature

internal_temperature:long_name Internal temperature
internal_humidity internal_humidity:units RH

Optics unit - circuit board humidity

internal_humidity:long_name Internal humidity
internal_humidity:comment Percent (%RH)
internal_pressure internal_pressure:units

hPa

Optics unit - circuit board pressure

internal_pressure:long_name Internal pressure
laser_temperature laser_temperature:units

Celsius

Temperature of the laser chip, measured inside the Peltier cooled package of the laser diode

laser_temperature:long_name Laser temperature
window_blower window_blower:long_name

Window blower

Window blower on/off information

window_blower:comment
  • 1 = on
  • 0 = off
internal_heater internal_heater:long_name Internal heater
internal_heater:comment
  • 1 = on
  • 0 = off
window_blower_heater window_blower_heater:long_name

Window blower heater

Window blower heater on/off information

window_blower_heater:comment
  • 1 = on
  • 0 = off

transmitter_enclosure_

temperature

transmitter_enclosure_temperature:units

Celsius

Hot side temperature of the laser peltier

transmitter_enclosure_temperature:long_name Transmitter enclosure temperature
Table 4. Variables, group status
Variable Attribute Description
time time:units

Seconds since 1970-01-01 00:00:00.000

The timestamp of the reported statuses. Common reporting frequency with profiles.

time:long_name Time
time:axis T
time:standard_name time
Device_controller_temperature Device_controller_temperature:long_name Controller unit CLC temperature
Device_controller_electronics Device_controller_electronics:long_name Controller unit CLC electronics
Device_controller_overall Device_controller_overall:long_name Controller unit CLC overall, the most severe status of CLC statuses
Optics_unit_accelerometer Optics_unit_accelerometer:long_name Optics unit CLO accelerometer
Optics_unit_electronics Optics_unit_electronics:long_name Optics unit CLO electronics
Optics_unit_overall Optics_unit_overall:long_name Optics unit CLO overall, the most severe status of CLO statuses
Optics_unit_memory Optics_unit_memory:long_name Optics unit CLO parameter memory
Optics_unit_tilt_angle Optics_unit_tilt_angle:long_name Optics unit CLO tilt angle
Receiver_electronics Receiver_electronics:long_name Receiver CLR electronics
Receiver_overall Receiver_overall:long_name Receiver CLR overall, the most severe status of CLR statuses
Receiver_memory Receiver_memory:long_name Receiver CLR parameter memory
Receiver_voltage Receiver_voltage:long_name Receiver voltage
Receiver_solar_saturation Receiver_solar_saturation:long_name Receiver solar saturation
Window_blocking Window_blocking:long_name Window blocking
Window_condition Window_condition:long_name Window condition
Window_blower_fan Window_blower_fan:long_name Window blower fan
Window_blower_heater Window_blower_heater:long_name Window blower heater
Servo_drive_electronics Servo_drive_electronics:long_name Servo drive CLS electronics
Servo_drive_overall Servo_drive_overall:long_name Servo drive CLS overall, the most severe status of CLS statuses
Servo_drive_memory Servo_drive_memory:long_name Servo drive CLS parameter memory
Servo_drive_control Servo_drive_control:long_name Servo drive CLS control
Servo_drive_ready Servo_drive_ready:long_name Servo drive CLS ready
Transmitter_electronics Transmitter_electronics:long_name Transmitter CLT electronics
Transmitter_light_source Transmitter_light_source:long_name Transmitter CLT light source
Transmitter_light_source_power Transmitter_light_source_power:long_name Transmitter CLT light source power
Transmitter_overall Transmitter_overall:long_name Transmitter CLT overall, the most severe status of CLT statuses
Transmitter_light_source_safety Transmitter_light_source_safety:long_name Transmitter light source safety
Transmitter_memory Transmitter_memory:long_name Transmitter CLT parameter memory
Maintenance_overall Maintenance_overall:long_name Device maintenance need
Device_overall Device_overall:long_name The most severe of all statuses
Recently_started Recently_started:long_name Recently started
Measurement_status Measurement_status:long_name Measurement status
Datacom_overall Datacom_overall:long_name Data communication status

Measurement_data_

destination_not_set

Measurement_data_destination_not_set:long_name Is the data destination set
Inside_heater Inside_heater:long_name Inside heater
Receiver_sensitivity Receiver_sensitivity:long_name Too big of a decrease signals receiver failure
Data_generation_status Data_generation_status:long_name Data generation status