The following tables show the parameters that are used with the get, getset, and set commands.
The following levels are in use:
- Level 0: read access
- Level 1: write access to basic parameters
- Level 2: write access to advanced parameters
To change the user level, use the level command.
Columns get and set indicate the required user level to view or change the setting.
Parameter | Description | Value / Range | get | set |
---|---|---|---|---|
cl.product.model | Product model | CL61D | 0 | − |
cl.product.modelname | Product model name | CL61 with Depolarization | 0 | − |
cl.product.revision | Product revision | A - Z | 0 | − |
Parameter | Description | Value / Range | get | set |
---|---|---|---|---|
cl.unit.alias1 | Alias name 1 | Freeform text | 0 | 1 |
cl.unit.latitude |
Latitude Unit: degrees |
−90.000000 … 90.000000 | 0 | 1 |
cl.unit.longitude |
Longitude Unit: degrees |
−180.000000 ... 180.000000 | 0 | 1 |
cl.unit.altitude |
Altitude Unit: meters Resolution: full meters Use negative values (−) for values below sea level. |
−999 ... 9999 | 0 | 1 |
cl.unit.azimuthangle |
Measurement azimuth angle Reference direction: north Unit: degrees |
0 ... 360 | 0 | 1 |
Parameter | Description | Value / Range | get | set |
---|---|---|---|---|
cl.password.admin | Admin user password | String | − | 1 |
Parameter | Description | Value / Range | Default | get | set |
---|---|---|---|---|---|
eth0.address.mac | MAC address | − | Set at factory, different for each device | 0 | − |
eth0.address.dhcp | DHCP |
|
− | 0 | − |
eth0.address.ip1 | IP address | 172.17.0.2 | − | 0 | − |
eth0.address.nm1 | Netmask | 255.255.255.0 | − | 0 | − |
eth0.address.gw1 | Gateway | 172.17.0.1 | − | 0 | − |
eth0.address.dns1 | DNS server1 | 172.17.0.1 | − | 0 | − |
eth0.address.dns2 | DNS server 2 | OFF |
− | 0 | − |
eth1.address.mac | MAC address | − | − | 0 | − |
eth1.address.dhcp | DHCP |
|
OFF |
0 | 1 |
eth1.address.ip1 | IP address | − | 192.168.2.102 | 0 | 1 |
eth1.address.nm1 | Netmask | − | 255.255.255.0 | 0 | 1 |
eth1.address.gw1 | Gateway | − | 192.168.2.1 | 0 | 1 |
eth1.address.dns1 | DNS server1 | − | 192.168.2.1 | 0 | 1 |
eth1.address.dns2 | DNS server2 |
|
OFF |
0 | 1 |
Restart the device after changing DNS settings. |
Parameter | Description | Value / Range | Default | get | set |
---|---|---|---|---|---|
time.sync.ntp1 | Network time protocol (NTP) server address 1 |
|
2.openembedded.pool.ntp.org | 0 | 1 |
time.sync.ntp2 | Network time protocol (NTP) server address 2 |
|
OFF |
0 | 1 |
time.unit.optime |
Unit operation since factory reset Unit: hours |
0 … 1000000 | − | 0 | − |
time.unit.now | Unit date and time, UTC | Format: YYYY-MM-DD HH:MM:SS |
− | 0 | 1 |
Parameter | Description | Value / Range | Default | get | set |
---|---|---|---|---|---|
algo.l1averaging. interval |
Number of 5‑second profiles averaged in each x‑polarization (XPOL), p‑polarization (PPOL), and attenuated backscatter profile (ABS). Averaging refers to the running average, so for every 5‑second profile, this parameter averages the number of previous profiles. The algorithm produces profiles in 5‑second intervals. You can specify that profiles recorded in netCDF files are averaged by n × 5 seconds, where n = 1 … 24. |
1 … 24 |
12 |
0 | 1 |
algo.l1averaging. depolinterval |
Number of 5‑second profiles averaged in each depolarization profile. For every 5-second profile, this parameter averages the number of previous profiles. The algorithm produces profiles in 5-second intervals. |
2 … 24 |
12 |
0 | 1 |
algo.fivelayerskycondition.enabled |
Default is When |
|
OFF |
0 | 1 |
algo.tiltanglecorrection. enabled |
Correction of cloud height by tilt angle. |
|
OFF |
0 | 1 |
algo.cloudlayers. offsetinmeters |
Corrected cloud height by offset value. For use when the device is installed above or below reference height. |
−999 ... 9999 | 0.0 | 0 | 1 |
algo.airplanefilter. maxrangeinmeters |
Filtering outliers caused by low-altitude airplanes from cloud data. For use when the device is installed at an airport. A typical maximum range for the filter is 400 m when the device is close to the runway. Unit: meters |
0 ... 1000 | 0 | 0 | 1 |
Parameter | Description | Value / Range | Default | get | set |
---|---|---|---|---|---|
reporting.destination. address |
Destination | Valid URL | − | 0 | 1 |
reporting.netcdf. institution |
Institution attribute in the .nc file | Freeform text | − | 0 | 1 |
reporting.netcdf. conventions |
NetCDF convention attribute in the .nc file https://cfconventions.org/Data/cf-conventions/cf-conventions-1.8/cf-conventions.html |
Acc. to CF-1.8 | − | 0 | − |
reporting.netcdf. comment |
Comment attribute in the .nc file | Freeform text | − | 0 | 1 |
reporting.netcdf.unit |
Unit used for reporting vertical visibility, cloud base heights, and range |
|
m | 0 | 1 |
reporting.profile. interval |
Interval for reporting to write out profiles. Only every nth profile is written to the netCDF file where the n is this parameter. This parameter can be used to avoid overlapping averaging. For example, if algo.l1averaging.interval is set to 6 corresponding to a 30‑second time span, this parameter can be set to 6 as well to get the profiles every 30 seconds. |
1 ... no maximum |
12 |
0 | 1 |
reporting.livefile. profilesperfile |
Number of profiles that are written in each .nc file. If reporting.profile.interval is set to 6 and this parameter is set to 4, one netCDF file will cover 2 min of data (6 × 4 × 5 s). |
1 … 1440 |
5 |
0 | 1 |
reporting.livefile. nameprefix |
Prefix in netCDF file name. See also cl.sender.fileprefix. NetCDF filename format: <cl.sender.fileprefix><reporting.livefile.nameprefix>_YYYYMMDD_HHmmss.nc |
Freeform text | − | 0 | 1 |
Parameter | Description | Value / Range | get | set |
---|---|---|---|---|
cl.sender.fileprefix |
Prefix in netCDF file name. Used to distinguish files coming from different CL61 devices. See also reporting.livefile.nameprefix and reporting.arhivefile.nameprefix. NetCDF filename format: <cl.sender.fileprefix><reporting.livefile.nameprefix>_YYYYMMDD_HHmmss.nc |
Freeform text | 0 | 1 |
cl.sender.hostpubmd5 | MD5 hash of the SSH server host key | String | 1 | 1 |
cl.sender.publickey | Public key | String | 1 | − |
Parameter | Description | Value / Range | Default | get | set |
---|---|---|---|---|---|
clt.factory. laserpower |
Transmitter laser power measured at the factory | − | Device specific | 2 | − |
clo.factory. cloudfactor |
Cloud calibration factor set at the factory | − | Device specific | 2 | − |
clo.field. cloudfactor |
Cloud calibration factor set by the user. The default is the same as clo.factory. cloudfactor |
− | Device specific | 2 | 2 |
Parameter | Description | Value / Range | Default | get | set |
---|---|---|---|---|---|
alerts.cl. recentlystarted |
After reset, time period for keeping recently restarted indication alert on. Unit: seconds |
0 … 3600 | 180 | 2 | − |
alerts.clo.tiltanglew |
Angle of ceilometer from vertical that triggers tilt angle warning Unit: degrees |
− | 45 | 2 | − |
alerts.clr. windowconditioni |
Limit for triggering window condition indication | − | 80 | 2 | − |
alerts.clr. windowconditionw |
Limit for triggering window condition warning | − | 50 | 2 | − |
alerts.clr. windowconditiona |
Limit for triggering window condition alarm | − | 20 | 2 | − |
alerts.clr. windowblockingratio |
Limit for triggering window blocking warning | − | 0.2 | 2 | − |
alerts.clt. lightsourcepowerw |
Limit for triggering TX light source power warning | − | 80 | 2 | − |
alerts.clt. lightsourcepowera |
Limit for triggering TX light source power alarm | − | 60 | 2 | − |
alerts.clt. lightsourcefailure |
Limit for triggering transmitter laser power failure | − | 10 | 2 | − |