Property | Description/Value |
---|---|
Measurement range | −70 … +180 °C (−94 … +356 °F) |
Sensor | Pt100 RTD Class F0.1 IEC 60751 |
Standard calibration 1 2 | |
Accuracy at +23 °C (+73.4 °F) |
±0.1 °C (±0.18 °F) |
Factory calibration uncertainty 3 | ±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F) |
Optional ISO 17025 calibration 4 2 | |
Accuracy at +23 °C (+73.4 °F) 1 | ±0.06 °C (±0.108 °F) |
Calibration uncertainty 3 | ±0.03 °C (±0.054 °F) |
Property | Description/Value |
---|---|
Operating temperature of probe body | −40 … +80 °C (−40 … +176 °F) |
Operating temperature of probe head | −70 … +180 °C (−94 … +356 °F) |
Storage temperature | −40 … +80 °C (−40 … +176 °F) |
Operating environment | Suitable for outdoor use |
IP rating | |
Probe body |
IP66: Dust-tight. Protected from powerful water jets from any direction. |
Probe head and cable |
IPX8/IPX9: Protected against high pressure and
temperature water jets. Protected from the effects of continuous immersion in
water. Tested under the following conditions:
|
Property | Description/Value |
---|---|
Operating voltage | 15–30 V DC |
Current consumption |
10 mA typical |
Digital output | RS-485, non-isolated |
Protocols | Modbus RTU |
Output parameters |
Temperature (°C) Water vapor saturation pressure (hPa) |
Property | Value/Description |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU) amended by 2015/863 |
Electromagnetic compatibility (EMC) |
EN 61326‑1, industrial environment |
Type approvals |
DNV GL certificate no. TAA00002YT |
Compliance marks |
CE, China RoHS, RCM |
Property | Description/Value |
---|---|
Connector | M12 5-pin A-coded male |
Weight (with a 2-m cable) | 224 g (7.9 oz) |
Probe cable length | 2 m (6.56 ft) or 10 m (32.8 ft) |
Materials | |
Probe | AISI 316L |
Probe body | AISI 316L |
Cable jacket | FEP |
1 Defined against calibration
reference. Including non-linearity, hysteresis, and repeatability.
2 Exposing temperature sensor to
temperatures outside −20 … +150 °C (−4 … +302 °F) may cause permanent additional
deviation of ±0.1 °C (0.18 °F)
3 Defined as ±2
standard deviation limits. Small variations possible; see calibration
certificate.
4 Accuracy depends on selected
calibration points. Accuracy with ISO 17025 calibration is defined here using a
5-point calibration in the following points: −30, −10, 0, +30, and +60 °C. For
more information on calibration services offered by Vaisala, see vaisala.com/calibration.