TMP1 specifications - HMP1 - HMP3 - HMP4 - HMP5 - HMP7 - HMP8 - HMP9 - HMPX - MMP8 - MMPX - TMP1 - TMPX

HMP Series with MMP8 and TMP1 User Guide

Document code
M212022EN
Revision
K
Language
English
Product
HMP1
HMP3
HMP4
HMP5
HMP7
HMP8
HMP9
HMPX
MMP8
MMPX
TMP1
TMPX
Document type
User guide
Table 1. TMP1 measurement performance
Property Description/Value
Measurement range −70 … +180 °C (−94 … +356 °F)
Sensor Pt100 RTD Class F0.1 IEC 60751
Standard calibration 1 2

Accuracy at +23 °C (+73.4 °F)

±0.1 °C (±0.18 °F)
Factory calibration uncertainty 3 ±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
Optional ISO 17025 calibration 4 2
Accuracy at +23 °C (+73.4 °F) 1 ±0.06 °C (±0.108 °F)
Calibration uncertainty 3 ±0.03 °C (±0.054 °F)
Figure 1. TMP1 temperature measurement accuracy over full range
Table 2. TMP1 operating environment
Property Description/Value
Operating temperature of probe body −40 … +80 °C (−40 … +176 °F)
Operating temperature of probe head −70 … +180 °C (−94 … +356 °F)
Storage temperature −40 … +80 °C (−40 … +176 °F)
Operating environment Suitable for outdoor use
IP rating
Probe body

IP66: Dust-tight. Protected from powerful water jets from any direction.

Probe head and cable
IPX8/IPX9: Protected against high pressure and temperature water jets. Protected from the effects of continuous immersion in water. Tested under the following conditions:
  • Immersion depth: 150 cm (59 in)
  • Immersion time: 60 min
Table 3. TMP1 inputs and outputs
Property Description/Value
Operating voltage 15–30 V DC
Current consumption

10 mA typical

Digital output RS-485, non-isolated
Protocols Modbus RTU
Output parameters

Temperature (°C)

Water vapor saturation pressure (hPa)

Table 4. TMP1 compliance
Property Value/Description

EU directives and regulations

EMC Directive (2014/30/EU)

RoHS Directive (2011/65/EU) amended by 2015/863

Electromagnetic compatibility (EMC)

EN 61326‑1, industrial environment

Type approvals

DNV GL certificate no. TAA00002YT

Compliance marks

CE, China RoHS, RCM

Table 5. TMP1 mechanical specifications
Property Description/Value
Connector M12 5-pin A-coded male
Weight (with a 2-m cable) 224 g (7.9 oz)
Probe cable length 2 m (6.56 ft) or 10 m (32.8 ft)
Materials
Probe AISI 316L
Probe body AISI 316L
Cable jacket FEP
Figure 2. TMP1 probe dimensions
1 Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2 Exposing temperature sensor to temperatures outside −20 … +150 °C (−4 … +302 °F) may cause permanent additional deviation of ±0.1 °C (0.18 °F)
3 Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
4 Accuracy depends on selected calibration points. Accuracy with ISO 17025 calibration is defined here using a 5-point calibration in the following points: −30, −10, 0, +30, and +60  °C. For more information on calibration services offered by Vaisala, see vaisala.com/calibration.