When you connect the probe to an Indigo200 or Indigo300 transmitter for the first time, the following analog output values per output parameter are active by default, unless you have ordered the transmitter with preconfigured analog outputs.
Output parameter | Output unit | AOUT min | AOUT max |
---|---|---|---|
Absolute humidity | g/m3 | 0 | 600 |
Absolute humidity at NTP | g/m3 | 0 | 600 |
Dew/frost point temperature | °C | −80 | 100 |
Dew/frost point temperature at 1 atm | °C | −80 | 100 |
Dew point temperature | °C | −80 | 100 |
Dew point temperature at 1 atm | °C | −80 | 100 |
Dew point temperature difference | °C | −10 | 50 |
Enthalpy | kJ/kg | −40 | 1500 |
Mixing ratio | g/kg | 0 | 500 |
Relative humidity | %RH | 0 | 100 |
Relative humidity (dew/frost) | %RH | 0 | 100 |
Relative saturation | %RS | 0 | 100 |
Temperature | °C | −40 | 60 |
Water activity | - | 0 | 1 |
Water concentration | ppmv | 0 | 1 000 000 |
Water concentration in oil | ppmw | 0 | 1 000 000 |
Water concentration (wet basis) | vol-% | 0 | 100 |
Water mass fraction | ppmw | 0 | 1 000 000 |
Wet-bulb temperature | °C | 0 | 100 |
Water vapor pressure | hPa | 0 | 1000 |
Water vapor saturation pressure | hPa | 0 | 1000 |