Example installation of a heated probe head - HMP1 - HMP3 - HMP4 - HMP5 - HMP7 - HMP8 - HMP9 - HMPX - MMP8 - MMPX - TMP1 - TMPX

HMP Series with MMP8 and TMP1 User Guide

Document code
M212022EN
Revision
K
Language
English
Product
HMP1
HMP3
HMP4
HMP5
HMP7
HMP8
HMP9
HMPX
MMP8
MMPX
TMP1
TMPX
Document type
User guide
Figure 1. Heated probe and temperature probe installation in a climate chamber
1
Temperature probe head
2
Humidity probe head with probe heating
3
Stainless steel cable tie or similar fastener
4
To be sealed, for example with silicone
5
Cable gland, for example: AGRO 1100.12.91.065
6
PTFE sleeve
7
Cable gland AGRO (Vaisala item code HMP247CG)
Let the probe cable hang loosely to prevent condensed water from running to the probe head. Avoid attaching the heated probe head in a way that conducts the heat away, as this may prevent the probe head from maintaining a temperature that is above dew point.