Property | Specification |
---|---|
Humidity sensor options | |
HUMICAP® R2 | Latest generation industrial sensor with improved corrosion resistance |
HUMICAP® 180V | Humidity sensor with a catalytic surface for processes with H2O2 |
Measurement range | 0 … 100 %RH |
Stability | ±0.5 %RH/year in typical HVAC applications |
Accuracy at 0 … +40 °C (+32 … +104 °F) 1 | |
0 … 90 %RH | ±1.5 %RH |
90 … 100 %RH | ±2.5 %RH |
Accuracy at +40 … +80 °C (+104 … +176 °F) and −40 … 0 °C (−40 … +32 °F) 1 2 | |
0 … 90 %RH | ±2.5 %RH |
90 … 100 %RH | ±3.5 %RH |
Factory calibration uncertainty | ±1.0 %RH |
Start-up and response time | |
Start-up time at +20 °C (+68 °F) | 8 s |
Response time (T63) at +20 °C (+68 °F) |
15 s |
Calculated humidity parameters (default analog output scale) | |
Dew point | −40 … +80 °C (−40 … +176 °F) |
Dew point / frost point | −40 … +80 °C (−40 … +176 °F) |
Absolute humidity | 0 … 300 g/m3 (0 … 131.1 gr/ft3) |
Wet bulb temperature | −40 … +80 °C (−40 … +176 °F) |
Enthalpy | −40 … 1600 kJ/kg (−9.5 … 695.6 Btu/lb) |
Mixing ratio | 0 … 600 g/kg (0 … 4200 gr/lb) |
Property | Specification |
---|---|
Temperature sensor | Pt1000 RTD Class F 0.1 IEC 60751 |
Measurement range | −40 … +80 °C (−40 … +176 °F) |
Default analog output scale | −20 … +80 °C (−4 … +176 °F) |
Accuracy at +20 °C (+68 °F) | ±0.1 °C (0.18 °F) |
Temperature dependence | ±0.005 °C/°C |
Factory calibration uncertainty | ±0.1 °C (0.18 °F) |
Response time (T63) with free convection | 8 min |
Property | Specification |
---|---|
Operating temperature, electronics | −40 … +60 °C (−40 … +140 °F) |
Operating temperature, probe | −40 … +80 °C (−40 … +176 °F) |
Storage temperature range | −40 … +80 °C (−40 … +176 °F) |
Maximum flow speed | 50 m/s with sintered filter |
Electromagnetic compatibility | EN61326-1, Industrial Environment |
Property | Specification |
---|---|
Power supply input |
10 … 35 VDC (RL = 0 Ω) 20 … 35 VDC (RL = 600 Ω) |
Analog output |
TMD62: 1 × T output 4 … 20 mA HMD62: 1 × RH output 4 … 20 mA, 1 × T output 4 … 20 mA 3 |
Service port | M8 4-pin male connector:
|
Screw terminal wire size | 0.5 … 2.5 mm2 |
Property | Specification |
---|---|
Housing material | Cast aluminum |
Probe material | Stainless steel |
IP rating | IP66 (NEMA 4X) |
Weight | 511 g (18 oz) |
1 Including non-linearity,
hysteresis, and repeatability
2 With HUMICAP® 180V sensor, accuracy is not specified below −20 °C (−4 °F)
operating temperature
3 Calculated output parameters for HMD62 and HMD65 include Td,
Tdf, A, X, Tw, and H.
4 Vaisala Insight software for Windows® available at www.vaisala.com/insight.