Calibration and adjustment with MI70 handheld indicator - HMD60 - HMD62 - MI70 - TMD60 - TMD62

HMD62 TMD62 User Guide

Document code
M212016EN
Revision
D
Language
English
Product
HMD60
HMD62
TMD60
TMD62
Document type
User guide
You can use the MI70 handheld indicator to make a 1-point adjustment to temperature (T) or relative humidity (RH). To connect the transmitter to MI70, you need the optional MI70 connection cable (Vaisala order code 219980SP). For instructions on connecting the transmitter to MI70 and general information about using MI70, see Overview of MI70 support.
Before adjusting the measurement with MI70, turn the adjustment trimmer(s) on the transmitter component board back to the ±0 position. When you adjust the measurement with MI70, the position in which the trimmer is at that moment becomes the new ±0 point.