To apply an accurate relative humidity compensation, the temperature and pressure compensation configurations must also match your measurement environment. See the rhcmode, tcmode and pcmode commands for instructions on enabling compensation configuration, and env command for instructions on setting a compensation value. |
For more information on environmental compensation and the default (neutral) compensation values used for disabled compensations, see Environmental Compensation on page 13 |
Syntax | Description | ||
---|---|---|---|
env<cr> | Show current compensation values. Before using this command, you must enable environmental compensation using the following commands:
|
||
env [temp | pres | oxy | hum] [value]<cr> | Set new permanent compensation values and store them in
eeprom. Eeprom:
temp = Compensation temperature. Range -40 … +100 °C. pres = Compensation pressure. Range 500 … 1100 hPa. oxy = Oxygen content of background gas. Range 0 … 100 %. hum = Relative humidity of background gas. Range 0 … 100 %. |
||
env [xtemp | xpres | xoxy | xhum] [value]<cr> | Set new compensation values and store them in RAM. RAM: n Volatile memory that loses the values when probe is reset, and where values are loaded from non-volatile memory at startup. n Must be used for continuously updated values. xtemp = Compensation temperature stored in RAM. Range -40 … 100 °C. xpres = Compensation pressure stored in RAM. Range 500 … 1100 hPa. xoxy = Oxygen content of background gas stored in RAM. Range 0 … 100 %. xhum = Relative humidity of background gas stored in RAM. Range 0 … 100 %.
|
||
Example (Show current compensation values;
all compensations are enabled. Note that temperature compensation is in "measured"
mode, so the value in use is constantly
changing):
|
|||
Example (set temperature compensation to
setpoint mode, and change temperature setpoint value to 5.00 °C in
RAM):
|
Syntax | Description |
---|---|
o2cmode<cr> | Check current oxygen compensation mode. Possible modes:
|
o2cmode [on | off]<cr> | Change oxygen compensation mode (on or off). |
Example (check oxygen compensation mode;
oxygen compensation is disabled, a neutral value is
used):
|
|
Example (enable oxygen compensation):
|
Syntax | Description |
---|---|
pcmode<cr> | Check current pressure compensation mode. Possible modes:
|
pcmode [on | off]<cr> | Change pressure compensation mode (on or off). |
Example (check pressure compensation mode;
pressure compensation is enabled using a setpoint
value):
|
Syntax | Description | ||
---|---|---|---|
rhcmode<cr> | Check current relative humidity compensation mode. Possible modes:
|
||
rhcmode [on | off]<cr> | Change relative humidity compensation mode (on or off). | ||
Example (check relative humidity
compensation mode; relative humidity compensation is disabled, a neutral value is
used):
|
|||
Example (enable temperature, pressure and
relative humidity compensation using setpoint values):
|
Syntax | Description |
---|---|
tcmode<cr> | Check current temperature compensation mode. Possible modes:
|
tcmode [on | off | measured]<cr> | Change temperature compensation mode (on, off or measured). |
Example (check temperature compensation
mode; temperature compensation is enabled using a setpoint value):
|
|
Example (change temperature compensation to
use internal measurement):
|