Wiring - GMP252

GMP252 User Guide

Document code
M211897EN
Revision
E
Language
English
Product
GMP252
Document type
User guide
Table 1. M12 male connector
Pin# Function Note Cable 223263SP wire colors
1 Power in
  • With digital output: 12 ... 30 V DC
  • With voltage output: 12 ... 30 V DC
  • With current output: 20 ... 30 V DC
Typical average power consumption <0.4 W, maximum 0.5 W.
Brown
2

RS-485-

or voltage output
Voltage:

0 ... 5 V DC or 0 ... 10 V DC (default analog output scaling)

White
3 GND Blue
4

RS-485 +

or current output
Current:

0 … 20 mA or 4 ... 20 mA (default analog output scaling)

Black
5 Output control

Connecting pin #5 to GND (pin #3) forces the probe to analog output mode. If an analog output configuration has not been selected, default 0 ... 10 V DC and 4 ... 20 mA scalings are used.

If pin #5 is not connected, the analog or digital output selected when ordering or set later through configuration is used.

Gray
Note that the probe always remains in analog mode when pin #5 is connected to pin #3, and cannot be switched to digital output in this wiring option.