The notation <cr> refers to the carriage return control character, which you can send in a terminal application by pressing enter on your keyboard. Before entering commands, send a <cr> to clear the command buffer.
You can enter the commands in uppercase or lowercase. In the command examples, the keyboard input by the user is in bold type.
Table 1 lists the basic serial commands that are available by default. To access advanced serial commands (listed in Table 2), enter the command pass 1300.
Command | Description |
---|---|
Device information and status | |
? | Show probe information. |
?? | Show probe information (will respond in POLL mode). |
errs | Show currently active errors. |
help | Show list of currently available serial commands. |
snum | Show probe serial number. |
system | Show probe firmware information. |
time | Show probe operation hours and uptime. |
vers | Show probe firmware version. |
Serial line output and communication | |
close | Close connection to probe (POLL mode) |
form [modifier string] | Show or set output format. |
intv [0 ... 255 s/min/h] | Set continuous output interval for R command. |
open [address] | Open connection to probe in POLL mode. |
r | Start the continuous outputting. |
s | Stop the continuous outputting. |
sdelay [1 ... 255] | Show or set serial line transmission delay (1 ... 255, 1 unit = 4 milliseconds) |
send | Output a single measurement message. |
seri [baud data stop parity] | Show or set the serial interface settings. |
smode [mode] | Show or set startup serial mode: RUN, STOP, or POLL. |
Environmental compensation | |
env | Show or set environmental parameters. |
Adjustment information | |
adate | Show CO2 factory adjustment date. |
atext | Show CO2 factory adjustment information. |
Other commands | |
reset | Reset the probe. |
pass [1300] | Access advanced serial commands. |
Command | Description |
---|---|
Serial line output and communication | |
addr [0 … 254] | Show or set probe address. |
Analog output | |
amode | Show or set analog output mode (analog output limits and error level). |
aover | Show or set analog output overrange and clipping behavior. |
asel | Show or set analog output parameter and scaling. |
Calibration and adjustment | |
cco2 | Adjust CO2 measurement gain and offset. |
cdate | Show or set calibration date. |
ct | Adjust temperature measurement offset. |
ctext | Show or set calibration information. |
Environmental compensation | |
o2cmode | Show or set oxygen compensation mode. |
pcmode | Show or set pressure compensation mode. |
rhcmode | Show or set humidity compensation mode. |
tcmode | Show or set temperature compensation mode. |
Other commands | |
frestore | Restore probe to factory settings. |