For optimal measurement accuracy, you can
use the Indigo80 handheld indicator to give additional information about the connected
probe’s measurement environment.
| The environment settings
available via Indigo80 correspond to the environmental compensation settings of the probe.
Therefore, Environment settings and are interconnected: the configuration set in either menu is applied to
both. |
-
Connect the probe(s) to the indicator.
-
Open the
main menu by pressing
.
-
Select Devices. If you have more than one device connected to the indicator, make a further selection
between the devices.
-
Select Environment settings to access the settings for your probe.
The available environment settings
depend on the probe and probe firmware version.
- Pressure
-
Temporary value for pressure
compensation. At device restart, compensation value changes back to power-up
default value.
- Temperature
-
Temporary value for
temperature compensation. At device reset, compensation value changes back to the
power-up default value.
- Relative humidity
-
Temporary value for relative
humidity compensation. At device reset, compensation value changes back to the
power-up default value.
- Oxygen concentration
-
Temporary value for oxygen
concentration compensation. At device reset, compensation value changes back to
the power-up default value.
-
Select a setting from the list, for example, Pressure.
-
Select the unit, if prompted.
-
Use the arrow buttons to set the desired value, and then select OK or Set.
You can also select to read the
value from another probe connected to the indicator, provided that the probe can
output the required compensation parameter. Press
to
highlight the Read value from field, and then press
to select the other probe as the source of the value.
-
Scroll down the Environment settings view and check that the compensation settings corresponding to the environment settings
you defined in the previous steps are set to On.
For example:
- If you set a value for Pressure, make sure that Pressure compensation on/off is set to On as well.
- If you set a value for temperature, make sure that Temperature is set to Setpoint.
-
Exit the menu
by pressing
or return to the main menu by pressing
.