Analog output configuration example - GMP252
GMP252 User Guide
- Document code
- M211897EN
- Revision
- E
- Language
- English
- Product
- GMP252
- Document type
- User guide
- 4 ... 20 mA current output
-
Connect the
transmitter to the Insight software.
-
Select
and open the Configure device menu.
-
Open the General menu and select Current output from the dropdown. Store your selection with Save.
-
Return to the Configure device menu and select one of the 3 analog outputs.
-
Select 4 ... 20 mA from the Output mode dropdown.
-
Type 0 in the Scale low end field.
-
Type 3.6 in the Error output level field.
-
Select Save to take your configuration into use.