Restoring factory adjustment with Indigo80 handheld indicator - GMP252

GMP252 User Guide

Document code
M211897EN
Revision
E
Language
English
Product
GMP252
Document type
User guide
  • Probe connection cable

If needed, you can restore the factory adjustment for your probe using the Indigo80 handheld indicator.

Always restore factory adjustment before entering a new adjustment. This prevents any possible earlier adjustments having an effect on the new adjustment you make.

  1. Connect the probe(s) to the indicator.
  2. Open the main menu by pressing .
  3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
  4. Select Calibrate / Adjust.
  5. Select any of the available parameters to access the view where you can restore the factory adjustment.
  6. Scroll down and select Restore factory adjustment > OK.
  7. Confirm by selecting OK.
  8. Press to end calibration.