Register number | Address | Description | Data format | Notes |
---|---|---|---|---|
2049 | 0800hex | Device status | 16-bit |
0 = Status OK. 1 = Critical error. 2 = Error. 4 = Warning. |
2050 | 0801hex | CO2 status | 16-bit |
0 = Status OK. 2 = CO2 reading not reliable. Appears during transmitter start-up. 256 = Measurement not ready. Appears during transmitter start-up. |
2052 | 0803hex | Error code | 32-bit field | See table Error codes in register 0803hex below. |
Bitmask | Error message | Severity |
---|---|---|
0000hex | Status OK. | |
0001hex | Program memory crc critical error. | Critical |
0002hex | Parameter memory crc critical error. | Critical |
0010hex | Low supply voltage error. | Error |
0020hex | Internal 30V error. | Error |
0040hex | Low RX signal error. | Error |
0080hex | Internal 8V error. | Error |
0100hex | RX signal cut error. | Error |
1000hex | Out of measurement range error. | Error |
2000hex | Sensor heater error. | Error |
4000hex | IR temperature error. | Error |
8000hex | FPI slope error. | Error |
10000hex | Internal 2.5V error. | Error |
20000hex | Internal 1.7V error. | Error |
40000hex | Low IR current error. | Error |
If several errors are active at the same time, the Modbus error code register contains the sum of currently active error codes.