Wiring through bottom lead-throughs - INDIGO300

Indigo300 User Guide

Document code
M212849EN
Revision
A
Language
English
Product
INDIGO300
Document type
User guide
  1. Before connecting wires or cables, make sure that the transmitter is powered off.
  2. Open the transmitter cover, if not already open.
  3. Open the cable gland and insert a suitable length of the cable inside the transmitter through the gland.
  4. Connect the wiring to the applicable screw terminals.
  5. Tighten the cable gland. Avoid overtightening.
    • Tightening torque for M20 gland: 8 Nm
    • Tightening torque for M16 gland: 6 Nm
  6. Reattach the transmitter cover.
    Tighten the screws hand-tight using the 4-mm Allen key (recommended tightening torque 3.6 Nm).
  7. Plug or seal the unused lead-throughs.