Analog output configuration example - INDIGO300

Indigo300 User Guide

Document code
M212849EN
Revision
A
Language
English
Product
INDIGO300
Document type
User guide

This example shows how to use Indigo80 to configure the transmitter's analog output 1 to use the following settings when a humidity and temperature probe is connected to the transmitter:

  • 4 ... 20 mA current output
  • Relative humidity measurement scaled to 0 ... 100 %RH
  • Clipping limit 5 %, error limit 10 %, and error output 3.6 mA
  1. Connect the transmitter to the indicator.
  2. Open the main menu by pressing .
  3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.

    In this case, select Indigo300.

  4. Select Settings.
  5. In Analog output type, select Current output.
  6. In Settings, scroll down and select Analog output 1.
  7. In Output mode, select 4...20 mA.
  8. In Output parameter, select Relative humidity.
  9. Select Scale low end. Use the arrow buttons to set the value as 0.00, and then select Set.
  10. Select Scale high end. Use the arrow buttons to set the value as 100.00, and then select Set.
  11. Select Error output level. Use the arrow buttons to set the value as 3.60, and then select Set.
  12. Select Clipping limit. Use the arrow buttons to set the value as 5.0, and then select Set.
  13. Select Error limit. Use the arrow buttons to set the value as 10.0, and then select Set.

    Your configuration is now in use.

  14. Exit the menu by pressing or return to the main menu by pressing .