Measurement type | Probe models |
---|---|
Humidity and temperature | HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9 |
Temperature | TMP1 |
Dew point | DMP5, DMP6, DMP7, DMP8 |
Carbon dioxide | GMP251, GMP252 |
Vaporized hydrogen peroxide | HPP271, HPP272 |
Moisture in oil | MMP8 |
Property | Description/Value |
---|---|
Operating temperature |
−20 … +60 °C (−4 …+140 °F) |
Storage temperature |
−30 … +70 °C (−22 …+158 °F) |
Operating humidity |
0 … 100 %RH |
Maximum operating altitude |
5000 m (approx. 16 400 ft) |
IP rating |
IP65: Dust-tight. Protected from powerful water jets from any direction. |
Note: Protect the device from direct sunlight. |
Property | Description/Value |
---|---|
Power supply input |
15 … 30 V DC 1 24 V AC ±10 % 50/60 Hz |
Fuse size for power supply | 2.5 A |
Transmitter service port connection |
|
Analog outputs | Current or voltage |
Number of analog outputs | 3 |
Selectable voltage output types | 0 … 1 V, 0 … 5 V, 0 … 10 V, 1 … 5 V, scalable, RL ≥ 10 kΩ |
Selectable current output types | 4 … 20 mA, 0 … 20 mA, scalable, RL ≤ 500 Ω |
Accuracy of analog outputs at 20 °C (+68 °F) | ±0.1 % full scale 3 |
Temperature dependence | ±0.005 % / °C full scale |
Current consumption at 20 °C (+68 °F) (Uin 24 V DC) | |
Minimum consumption with display off, no analog outputs active, no probe connected 4 | 13 mA |
Minimum consumption with display on, brightness normal mode | 31 mA |
Uout 0 … 1 V, 0 … 5 V, 0 … 10 V, 1 … 5 V | Max. + 1.8 mA per connected channel |
Iout 4 … 20 mA, 0 … 20 mA | Max. + 12.3 mA per connected channel |
Property | Description/Value |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU) amended by 2015/863 |
Electromagnetic compatibility (EMC) |
EN IEC 61326-1, industrial environment |
EMC emissions |
CISPR 32 / EN 55032, Class A FCC part 15 B, Class A ICES‑3 / NMB‑3 (Class A) |
Compliance marks |
CE, China RoHS, FCC, RCM, UKCA |
Property | Description/Value |
---|---|
Housing material | EN AW-6082 |
Connection screw terminals | Max. 1.5 mm2 wire (16 AWG) |
Cable lead-throughs for output and power cables |
|
Cable diameter for M20×1.5 gland | 7 … 13 mm (0.26 … 0.51 in) |
Cable diameter for M16×1.5 gland | 2 … 6 mm (0.08 … 0.24 in) |
Dimensions |
161 × 134 × 43.5 mm (6.34 × 5.26 × 1.71 in) |
Weight | 1200 g (2.65 lb) |
1 When used with the HMP7 probe, the minimum required power
supply input is 18 V DC.
2 Vaisala
Insight software for Windows® available at http://www.vaisala.com/insight.
3 For the voltage outputs, small
variation is possible around true zero.
4 For the current consumption of the connected
probe, see the probe's user documentation.