If you are using the connector plate, not all figures in Connecting cables and grounding apply. |
- 1
- Power connector
- 2
- RS‑485 signal connector
- 3
- Power connector slot +. Connect to + on the power cable.
- 4
- Power connector slot − (ground). Connect to − on the power cable.
- 5
- Signal connector slot 1. Connect to R/D− on the data cable.
- 6
- Signal connector slot 2. Connect to R/D+ on the data cable.
If the display is delivered with the optional connector plate, the DIP switches are set to the RS‑485 2‑wire mode already at the Vaisala factory. If you install the connector plate after the delivery, set the DIP switches to the RS‑485 2‑wire mode yourself. |