Parameters LIMIT and HYSTERESIS are used to switch the relay based on concentration or temperature measurement values by the functions CONC - HIGH LIMIT, CONC - LOW LIMIT, T - HIGH LIMIT and T - LOW LIMIT. The relay closes on high limit when the measured value exceeds HIGH LIMIT + HYSTERESIS. The relay opens on high limit when the measured value goes below HIGH LIMIT - HYSTERESIS. Alternatively, the relay closes on low limit when the measured value goes below LOW LIMIT - HYSTERESIS. The relay opens on low limit when the measured value exceeds LOW LIMIT + HYSTERESIS.