  1. To view the parameters, select Calibration > Summary.

    The settings in the upper part of the Calibration display are different types of lists that sets parameters for showing the measurement data, for example measurement units.

    In the lower part of the window, the words are links to the given settings. For example, by tapping Field, you get access to the field calibration parameters. If you want to see all the parameters, tap Summary.

    The summary extends to a second page, a small arrow on the top right takes you to the second (and third) page.
    1. Tap the little arrow in the top right to see page 2.
    2. Tap the little arrow in the top left to return to page 1.
  2. To adjust the parameters, do the following.
    1. Go to Calibration and choose the appropriate parameter set, for example Field.
    2. In the list, tap the parameter you want to change, type the new numbers with the number pad and approve the change by tapping DONE.
      The parameter lists for Chemical and Field do not fit on the screen. Swipe down or use the scroll bar on the right to view the bottommost parameters.