The MI implements an Ethernet / IP adapter that provides measurement data for connected scanners. Currently active MI app configuration determines the available measurement data and the Ethernet/IP connection parameters. Connection can be set up easily with the Windows PC tool “Vaisala MI/CI EDS Generator” that generates EDS files based on the requested MI app configuration. The tool is available at https://www.vaisala.com/en/industrial-measurements/products/liquid-concentration. When using the tool, select the apps that are configured for each appslot on MI, then press GENERATE EDS FILE and choose the name and location for the new EDS file.
The data content of the O > T and T > O instances is a combination of individual app data contents. The following tables provide the data content of individual apps.
Name | Bytes | Type | Function |
---|---|---|---|
Conc | 0-3 | REAL | Final concentration value |
Temp | 4-7 | REAL | Process temperature |
Calc | 8-11 | REAL | Calculated concentration value |
nD | 12-15 | REAL | Calculated refractive index value |
S/N | 16-27 | STRING | Refractometer serial number |
Status | 28-59 | STRING | Refractometer status message |
Name | Bytes | Type | Function |
---|---|---|---|
ExtWash | 0-1 | INT | External wash request input |
Name | Bytes | Type | Function |
---|---|---|---|
RecipeID | 0-1 | INT | Recipe identification number input |
The connection can be set up manually with the following parameters: Port: 0xAF12 (44818)
O > T:
- Instance number: 119
- Data size: sum of configured app’s input data sizes, see the tables above.
- Real time format: modeless
- Packet rate: 1000 ms
T > O:
- Instance number: 109
- Data size: sum of configured app’s output data sizes, see the tables above.
- Real time format: modeless
- Packet rate: 1000 ms
- Connection type: point to point