Configuring mA inputs - PR-43-AC - PR-43-AP - PR-43-GC - PR-43-GP - PR-43-PC
MI User Guide
- Document code
- IM-EN-MI
- Revision
- D
- Language
- English
- Product
- PR-43-AC
- PR-43-AP
- PR-43-GC
- PR-43-GP
- PR-43-PC
- Document type
- User guide
-
Select the MULTI application.
-
Set the function.
Choose between
COND for conductivity and
TEMP for temperature.
-
Set the parameters for the function.
- Minimum and maximum values set the measurement range.
- The 4 … 20 mA input range scales linearly between minimum and maximum value set by
parameters so that 4 mA input corresponds minimum value and 20 mA input corresponds
maximum value.
- Poll interval secs define the sample rate of mA input.
-
To exit the setup, tap Done.