Item | Specification |
---|---|
Display | xga 1024 × 768 pixels, touch screen |
Power | AC input 100 … 240 V AC, ±10 %, 50 / 60 Hz, 60 VA, optional 24 V DC |
Overvoltage category | OVC II |
Refractometer connectivity | 1 to 4 refractometers can be connected to the MI. Refractometers are independent of each other: own parameter sets and usable in different applications. |
Ethernet connection | 10 / 100 Mbit/s, data acquisition over UDP / IP Protocol |
Modules | 1 to 8 modules for added connectivity |
Apps | 1 to 4 apps |
IP rating |
Outer door closed: IP67: Dust-tight. Protected from the effects of temporary immersion in water under standardized conditions of pressure and time. Outer door open and display panel fixed: IP66: Dust-tight. Protected from powerful water jets from any direction. |
NEMA rating |
NEMA 4X: Dust-tight. Protected from corrosion and powerful water jets from any direction. |
Installation |
Indoor/outdoor, max. +50 °C (+122 °F), min. 0 °C (+32 °F) Must be protected from rain and direct sunlight. During installation highest relative humidity of ambient air 80 %. |
Pollution degree |
3 |
Maximum operating altitude |
2000 m (approx. 6500 ft) |
Weight | 12.9 kg (28.5 lbs) |
Regulatory compliance | |
EU directives and regulations |
RoHS Directive (2011/65/EU) amended by 2015/863 EMC Directive (2014/30/EU) Low Voltage Directive (2014/35/EU) |
Electrical safety |
IEC/EN/UL/CSA 61010‑1 |
EMC compatibility |
IEC/EN 61326-1, industrial environment IEC/EN 61000-3-2 IEC/EN 61000-3-3 |
Item | Specification |
---|---|
Relay module | 2 signal relays 150 V / 3 A |
mA output module | 2 × 4 … 10 mA, max. load 1000 Ω, galvanic isolation 1000 V DC |