  • Adjustable wrench
To maintain cabinet tightness and provide strain relief to the cables, all cable glands on the DGA monitor and the power supply unit must be tightened. Unused cable glands must remain plugged.
  1. Check every cable gland that is in use:
    1. Pull on the cable slightly to verify that the cable is securely held by the cable gland.
    2. Tighten the cable gland if the cable moves easily. See Figure 1 and Figure 1 for recommended tightening torques.
  2. Check that every unused cable gland is plugged and tightened.