PR-43-AC mounting specifications - PR-43-AC - PR-43-AP

PR-43-AC/AP User Guide

Document code
IM-EN-PR43ACAP
Revision
D
Language
English
Product
PR-43-AC
PR-43-AP
Document type
User guide

The sanitary process refractometer PR‑43‑AC is connected to the process by a 2.5 in sanitary clamp. The recommended mounting is in a pipe bend, with a vertical flow upwards before the refractometer, and a horizontal pipe after. This mounting enables the following:

  1. Self-cleaning of prism due to the flow directed against its surface.
  2. Efficient drainage when the pipe is emptied.

For pipe diameters of 80 mm (3 in) or bigger, a ferrule is welded directly to the pipe wall. A ferrule, length 21.5 mm (0.85 in), is delivered with standard refractometer delivery from Vaisala).

Figure 1. Mounting with sanitary ferrule, pipe diameter 80 mm (3 in) or more

Flow cells are available for smaller pipe diameters. The following figures show the flow cells for smaller pipe diameters. The flow cells are exchangeable with standard 90° bend pieces.

Figure 2. Flow cell AFC-HSS-H10 for pipe diameter 25 mm (1 in) and H15 for pipe diameter 40 mm (1.5 in)
Figure 3. Flow cell AFC-HSS- with wash nozzle connection (-NC) H10 for pipe diameter 25 mm (1 in) and H15 for pipe diameter 40 mm (1.5 in)
Flow cell type A B C D
AFC‑H10‑RI‑NC 16 mm (0.630 in) 22.6 mm (0.890 in) 25.6 mm (1.008 in) 50.8 mm (2 in)
AFC‑H10‑SI‑NC 22.6 mm (0.890 in) 22.6 mm (0.890 in) 25.6 mm (1.008 in) 50.8 mm (2 in)
AFC‑H15‑RI‑NC 22.6 mm (0.890 in) 35.6 mm (1.402 in) 38.6 mm (1.520 in) 70.3 mm (2.768 in)
AFC‑H15‑SI‑NC 35.6 mm (1.402 in) 35.6 mm (1.402 in) 38.6 mm (1.520 in) 70.3 mm (2.768 in)
Figure 4. Flow cell AFC-HSS-H20 for pipe diameter 50 mm (2 in) and H25 for pipe diameter 65 mm (2.5 in)
Figure 5. Flow cell AFC-HSS- with wash nozzle connection (-NC) H20 for pipe diameter 50 mm (2 in) and H25 for pipe diameter 65 mm (2.5 in)
Figure 6. Side flow cells