Prism wash systems - PR-43-AC - PR-43-AP

PR-43-AC/AP User Guide

Document code
IM-EN-PR43ACAP
Revision
D
Language
English
Product
PR-43-AC
PR-43-AP
Document type
User guide
Figure 1. Prism wash system for steam. The following figures describe the prism wash system for steam.
Table 1. Prism wash system for steam
No. Description Supplied by Qty
1 PR‑43 refractometer Vaisala 1
2 Multichannel user interface Vaisala 1
3 Safety switch PR‑7060 Vaisala 1
4 Platform 4 cable PR‑8430 Vaisala 1
5 Relay cable 2x1 (AWG 17) Customer 1
6 Power supply Customer 1
6.1 100‑240 V  AC/50‑60 Hz
6.2 24 V DC
7 Flexible steam pipe ¼ in x 24 in, PR‑3515 Vaisala 1
8 Shut‑off valve and steam trap PR‑3340‑230/110/24 V DC Vaisala 1
9 Strainer PR‑3342 Vaisala 1
10 Solenoid cable 3x1 (AWG 17) Customer 1
11 Mains power switch PR‑10900 Vaisala 1
12 mA output cable Customer 1
13 Ethernet cable for interfaces PR‑8440 Vaisala 1
14 Instrument air line Customer 1
15 Pressure reducer and gauge PR‑3341‑J Vaisala 1
Figure 2. Wiring for prism wash system for steam
WARNING In high pressure wash systems, pressure increase can occur in a closed pipe section when the high pressure pump is operated. It is recommended to mount a pressure relief valve in the pipe section. Make sure that relief pressure is according to pipe pressure rating.
Figure 3. Prism wash system for high pressure water. The following figures describe the prism wash system for high pressure water.
Table 2. Prism wash system for high pressure water
No. Description Supplied by Qty
1 PR‑43 refractometer Vaisala 1
2 Multichannel user interface Vaisala 1
3 Power relay unit PR‑3603‑300‑U/M‑230/110 Vaisala 1
4 Platform 4 cable PR‑8430 Vaisala 1
5 Relay cable 3x1 (AWG 17) Customer 1
6 Power supply for MI Customer 1
6.1 Power supply 100‑240 V AC/50-60 Hz
6.2 Power supply 24 V DC
7 Water supply line min. 12 mm (0.5 in) Customer 1
8 High pressure pump PR-3602-XXX Vaisala 1
9 Solenoid cable 3x1 (AWG 17) Customer 1
10 Power cable 4x2.5 (AWG 12) Customer 1
11 Power supply 400‑575 V AC/50‑60 Hz Customer 1
12 Mains power switch PR‑10900 Vaisala 1
13 mA output cable Customer 1
14 Ethernet cable for interface PR‑8840 Customer 1
Figure 4. Wiring for prism wash system for high pressure water