DM70 can be used for making the temperature adjustment of DMT340 series transmitters.
- Turn off DM70.
- Connect the cable 211339 between DMT340 (service port) and DM70 (connector I or II). Turn on both devices.
- Press the ADJ button on the DMT340 component board to enable the adjustment mode.
- From the list of parameters, select TI/II.
- At this point you need to select the adjustment method. If the reference probe is connected, three options are displayed: To same as TI/II (see To same as TI/II), 1-point adjustment (see 1-point adjustment) or 2-point adjustment (see 2-point adjustment).
- You are prompted to check the settings: make sure that the pressure units of DM70 and DMT340 are the same. If not, change the unit by pressing Unit on the corresponding row.