This procedure is one of the alternative options for completing the adjustment described in Temperature adjustment of DM70. |
- After selecting 2-point adjustment, press Select to continue.
- Set the probe to a lower reference temperature. You can follow the stabilization from the graphical display by pressing Graph. When the reading has stabilized, press Ready.
- Use the arrow buttons to give the lower reference temperature value and then press OK.
- Set the probe to a higher reference temperature. You can follow the stabilization from the graphical display by pressing Graph. When the reading has stabilized, press Ready.
- Use the arrow buttons to give the higher reference temperature value and then press OK.
-
Confirm the adjustment by pressing
Yes. If you press
No, you return to the adjustment display and no changes are made.
If the difference between the 2 references is less than 30 %, the adjustment cannot be done. - The adjustment is complete. To exit the adjustment mode, press Back.
- To return to the basic display, press Exit.