2-point adjustment - DM70

DM70 User Guide

Document code
M010091EN
Revision
G
Language
English
Product
DM70
Document type
User guide
This procedure is one of the alternative options for completing the adjustment described in Temperature adjustment of DM70.
  1. After selecting 2-point adjustment, press Select to continue.
  2. Set the probe to a lower reference temperature. You can follow the stabilization from the graphical display by pressing Graph. When the reading has stabilized, press Ready.
  3. Use the arrow buttons to give the lower reference temperature value and then press OK.
  4. Set the probe to a higher reference temperature. You can follow the stabilization from the graphical display by pressing Graph. When the reading has stabilized, press Ready.
  5. Use the arrow buttons to give the higher reference temperature value and then press OK.
  6. Confirm the adjustment by pressing Yes. If you press No, you return to the adjustment display and no changes are made.
    If the difference between the 2 references is less than 30 %, the adjustment cannot be done.
  7. The adjustment is complete. To exit the adjustment mode, press Back.
  8. To return to the basic display, press Exit.