Temperature adjustment of DM70 - DM70

DM70 User Guide

Document code
M010091EN
Revision
G
Language
English
Product
DM70
Document type
User guide

The temperature adjustment can be done using the 1-point or 2-point adjustment. The 1-point adjustment should be done within the temperature range in which the device is most often used.

First carry out steps 1 ... 4. In 5, choose one of the two alternative methods for completing the adjustment.

To start the adjustment:

  1. Press the Calibration button of the probe handle (see number 4 in Figure 1) using a tool with a thin, sharp point, for example, a small screwdriver. The indicator will switch into the adjustment mode. To confirm calibration, press OK.
  2. Select Td/f and press Select.
  3. Select T and press Select.
  4. To select the adjustment method, first press Adjust and then choose the adjustment method, either 1-point adjustment or 2-point adjustment.
  5. Depending on your selection, continue with the instructions in 1-point adjustment or 2-point adjustment.