The temperature adjustment can be done using the 1-point or 2-point adjustment. The 1-point adjustment should be done within the temperature range in which the device is most often used.
First carry out steps 1 ... 4. In 5, choose one of the two alternative methods for completing the adjustment. |
To start the adjustment:
- Press the Calibration button of the probe handle (see number 4 in Figure 1) using a tool with a thin, sharp point, for example, a small screwdriver. The indicator will switch into the adjustment mode. To confirm calibration, press OK.
- Select Td/f and press Select.
- Select T and press Select.
- To select the adjustment method, first press Adjust and then choose the adjustment method, either 1-point adjustment or 2-point adjustment.
- Depending on your selection, continue with the instructions in 1-point adjustment or 2-point adjustment.