Performing manual autocalibration - DM70

DM70 User Guide

Document code
M010091EN
Revision
G
Language
English
Product
DM70
Document type
User guide
To check that autocalibration has taken place for obtaining the most accurate measurement in a very dry environment, you can autocalibrate manually before measuring as follows:
  1. Open the menu by pressing Open.
  2. Select Functions and press .
  3. Select AutoCal and press Start.
  4. If you want to start autocalibration, press Yes. If the autocalibration conditions are not fulfilled, a note appears on the display informing that the calibration cannot be made or that a new attempt will be made later.
  5. When the autocalibration is completed, the display returns to the basic display automatically.
    If it is too humid, autocalibration will not start.
    1. Display during the adjustment period (< 15 seconds).
    2. Adjustment period is completed and stabilization period will start.
    3. After successful autocalibration, the corrected readings are shown during the stabilization period. If the autocalibration fails, no readings are shown, only lines are displayed.
    4. Display during the stabilization period (1 ... 2 minutes).
    5. Autocalibration is completed.