For the DMP74A probe, the wet point humidity reference must be > 20 %RH. For the DMP74B and DMP74C probes , the wet point humidity reference must be < 20 %RH. |
For DMP74A, humidity references of 0 % (for example nitrogen) and 30 ... 75 % are recommended. Make sure that the difference between the humidity references is more than 30 %.
For DMP74B and DMP74C, humidity references of 0 % (for example nitrogen) and 11 % are recommended.
To make a 2-point %RH adjustment:
1 If there is a temperature difference between the generated reference and the probe temperature,
relative humidity must be calculated to correspond to the temperature of the probe.