1-point adjustment - DM70

DM70 User Guide

Document code
M010091EN
Revision
G
Language
English
Product
DM70
Document type
User guide
This procedure is one of the alternative options for completing the adjustment described in Temperature adjustment of DM70.
  1. After selecting 1-point adjustment, press Select to continue.
  2. Set the probe to the reference temperature. You can follow the stabilization from the graph display by pressing Graph. The minimum recommended stabilization time for this critical adjustment is 5 hours. When the reading has stabilized, press Ready.
  3. Use the arrow buttons to give the reference temperature and then press OK.
  4. Confirm the adjustment by pressing Yes. If you press No, you return to the adjustment mode display and no changes are made.
  5. To exit the adjustment mode, press Back.
  6. To return to the basic display, press Exit.