This procedure is one of the alternative options for completing the adjustment described in Temperature adjustment of DM70. |
- After selecting 1-point adjustment, press Select to continue.
- Set the probe to the reference temperature. You can follow the stabilization from the graph display by pressing Graph. The minimum recommended stabilization time for this critical adjustment is 5 hours. When the reading has stabilized, press Ready.
- Use the arrow buttons to give the reference temperature and then press OK.
- Confirm the adjustment by pressing Yes. If you press No, you return to the adjustment mode display and no changes are made.
- To exit the adjustment mode, press Back.
- To return to the basic display, press Exit.