This procedure is 1 of the alternative options for completing the adjustment described in Temperature adjustment of DMT340 series using DM70. |
- After selecting 2-point adjustment, press Select to continue.
- Set the probe to the lower reference temperature. You can follow the stabilization from the graphic display by pressing Graph. When the reading has stabilized, press Ready.
- Use the arrow buttons to give the lower reference temperature value and press OK.
- Set the probe to the higher reference temperature. You can follow the stabilization from the graphic display by pressing Graph. When the reading has stabilized, press Ready.
- Use the arrow buttons to give the higher reference temperature value and press OK.
-
Confirm the adjustment by pressing
Yes. If you press
No, you return to the adjustment mode display and no changes are
made.
If the difference between the two references is less than 30 °C, the adjustment cannot be done. - The adjustment is now complete. Press Back to exit the adjustment mode.
- To return to the basic display, press Exit.