Td/f adjustment of DM70 - DM70

DM70 User Guide

Document code
M010091EN
Revision
G
Language
English
Product
DM70
Document type
User guide
For DMP74B/DMP74C, the sensor purge must be performed manually 1 hour before adjusting the dew point.

For DMP74A, the reference temperature of the dew point should be −40 ˚C ... −60 ˚C (−40 ˚F ... −76 ˚F).

For DMP74B/DMP74C, the reference temperature of the dew point should be −57 ˚C ... −67 ˚C (−70.6 ˚F ... −88.6 ˚F).

To adjust the dew point:

  1. Set the probe to the reference condition. Follow the stabilization from the graphic display by pressing Graph. The minimum recommended stabilization time for this critical adjustment is 5 hours.
  2. Perform the manual purge if necessary.
  3. Press the Calibration button on the probe handle (see Figure 1) using, for example, a small screwdriver. The indicator will switch into adjustment mode. To confirm calibration, press OK.
  4. Select Td/f and press Select.
  5. To check the environmental settings, press Yes. To go to calibration directly, press No.
  6. To select 1-point adjustment, press Adjust and then Select.
  7. When the reading has stabilized at the reference level, press Ready.
  8. Use the arrow buttons to give the reference value. Then press OK.
  9. Confirm the adjustment by pressing Yes. If you press No, you return to the adjustment mode display and no changes are made.
  10. Wait until the automatic autocalibration has taken place.
  11. When the calibration completes, press Back to exit the adjustment mode.
  12. To return to the basic display, press Exit.
  13. Place the screw back onto the calibration button.
  14. To seal the calibration, attach a sticker on the calibration button.
    Several autocalibrations may be necessary after this adjustment has been performed until the transmitter reaches full accuracy.