DM70 and DMP74 specifications - DM70

DM70 User Guide

Document code
M010091EN
Revision
G
Language
English
Product
DM70
Document type
User guide
Table 1. DM70 handheld meter general specifications
Property Description/value
Storage temperature −40 … +70 ˚C (−40 … +158 ˚F)
Storage humidity range 0 … 100 % non-condensing
Housing material ABS/PC blend
Total weight 750 g (26 oz)
Electromagnetic compatibility

Complies with the following standard:

EN 61326-1:1997 +Am 1:1998, Electrical equipment for measurement, control, and laboratory use - EMC requirements: Basic environment.

Table 2. DMP74A, DMP74B, and DMP74C measurement performance
Property Description/value
Dew point temperature
Measurement range

DMP74C: −70 … +30 ˚C (−94 … +86 ˚F)

DMP74A: −50 … +60 ˚C (−58 … +140 ˚F)

DMP74B: −70 … +30 ˚C (−94 … +86 ˚F)

Accuracy ±2 ˚C (±3.6 ˚F), see the accuracy graphs below
Dew point range for ±2 ˚C. Accuracy for pressure converted dew point (dew point measured in pressure, calculated to 1 atm dew point value)

DMP74A: −64 … +60 ˚C (−83 … +140 ˚F)

DMP74B: −80 … +20 ˚C (−112 … +68 ˚F)

DMP74C: −72 … +20 ˚C (−98 … +68 ˚F)

DMP74A:
Response time: 63 % (90 %) at flow rate 0.2 m/s, 1 bar absolute pressure, +20 ˚C
0 … −40 ˚C Td (+32 … +40 ˚F Td) 20 s (120 s)
−40 … ˚C Td (−40 … +32 ˚F Td) 10 s (20 s)
DMP74B/DMP74C:
Response time: 63 % (90 %) at flow rate 0.2 m/s, 1 bar absolute pressure, +20 ˚C
0 … −60 ˚C Td (+32 … −76 ˚F Td) 50 s (340 s)
−60 … 0 ˚C Td (−76 … +32 ˚F Td) 10 s (20 s)
Temperature
Measurement range −10 … 60 ˚C (14 … 140 ˚F)
Accuracy at +20 ˚C (+68 ˚F) ±0.2 ˚C (±0.36 ˚F)
Typical temperature dependence of electronics ±0.005 ˚C/˚C (±0.005 °F/°F)
Temperature sensor Pt100 IEC751 1/3 class B
PPM (ppmv or ppmw)
Measurement range (typical)

DMP74A: 40 … 200000 ppm

DMP74B/DMP74C: 10 … 20000 ppm

Accuracy at +20 ˚C (+68 ˚F):
< 40 ppm ±(0.5 ppm + 25.4 % of reading)
> 40 ppm ±(7.3 ppm + 8.3 % of reading)
Absolute humidity (DMP74A probe recommended)
Measurement range (typical) 0.5 … 100 g/m3 (0.2 … 40 gr/ft3)
Accuracy

±(0.2 g/m3 + 10 % of reading)

±(0.1 gr/ft3 + 10 % of reading)

Mixing ratio (DMP74A probe recommended)
Measurement range (typical) 0.2 … 100 g/kg (2 … 700 gr/lbs)
Accuracy

±(0.1 g/kg + 12 % of reading)

±(0.8 gr/lbs + 12 % of reading)

Relative humidity (DMP74A)
Measurement range 0 … 100 %
Accuracy at +20 ˚C (+68 ˚F)

RH < 5 %: ±(0.025 % RH + 17.5 % of reading)

RH > 5 %: ±(0.8 % RH + 2 % of reading)

Measurement environment
Temperature −10 … +60 ˚C (+14 … +140 ˚F)
Pressure

DMP74A: 0 … 20 bara (0 … 290 psia)

DMP74B: 0 … 20 bara (0 … 290 psia)

DMP74C: 0 … 10 bara (0 … 150 psia)

Sample flow rate No effect on dew point accuracy
Figure 1. DMP74A dew point accuracy vs. measurement conditions
Figure 2. DMP74B and DMP74C dew point accuracy vs. measurement conditions

Dashed line: For DMP74C the ±2 °C accuracy range is limited to −50 °C Td when used in SF6 gas.

Table 3. DMP74A, DMP74B, and DMP74C general specifications
Property Description/value
Sensor

DMP74A: Vaisala DRYCAP® 180S

DMP74B/DMP74C: Vaisala DRYCAP® 180M

Probe material (wetted parts) Stainless steel (AISI 316L)
Sensor protection Sintered filter, (AISI 316L), item No. HM47280SP
Mechanical connection G1/2" ISO2281 thread with bonded seal ring (U-seal)
Housing classification IP65 (NEMA4)
Storage temperature −40 … +70 ˚C (−40 … +158 ˚F)
Storage humidity 0 … 100 %RH non-condensing
Weight 350 g (12 oz)
Table 4. DMP74A, DMP74B, and DMP74C mechanical specifications
Property Description/value
Materials
Probe housing material PET plastic
Filter PTFE membrane, PET plastic grid
Connector Nickel plated brass, M12 5-pin
Housing classification, probe body IP65
Connector M12 5-pin male
Weight, probe head 45 g (1.6 oz)
Dimensions
Probe diameter 25 mm (0.98 in)
Probe length 96 mm (3.8 in)
Table 5. Sampling cell specifications
Property Description/value
DMT242SC sampling cell
Inlet/outlet thread ISO G 3/8"/G 1/4"
Pressure limit 10 MPa, 100 barg, 1450 psig
Material Stainless steel AISI316
Weight 220 g (7.8 oz)
DMT242SC sampling cell with Swagelok connectors
Inlet/outlet thread 1/4" Swagelok connectors (for 1/4" tubing)
Pressure limit 4 MPa, 40 barg, 580 psig
Material Stainless steel AISI316
Weight 285 g (10 oz)
DSC74 sampling cell for pressurized gases
Quick connector Type D/NIP08
Leak screw Screwdriver operated
Inlet/outlet thread ISO G 3/8"/G 1/4"
Thread adapters included

A) ISO G 3/8" to G 1/4" (female)

B) ISO G 3/8" to G 1/2" (male)

Pressure limit 1 MPa, 10 barg, 145 psig
Material Stainless steel AISI316
Weight 300 g (11 oz)
DSC74B 2-pressure sampling cell
Inlet/outlet thread ISO G 3/8"/G 3/8"
Flow adjustment Manually operated
Pressure limit 1 MPa, 10 barg, 145 psig
Material Stainless steel AISI316
Weight 390 g (14 oz)
DSC74C
DSC74C is a sampling cell that combines DSC74B with DMCOIL cooling/venting coil.
DMCOIL
Coil pipe size 6 mm (0.24 in)
Connection to sampling cell With ISO G 1/4" and G 3/8" thread adapter
Weight 130 g (4.6 oz)
DSS70A sampling system
Measuring range −60 ˚C (−76 ˚F) Td … Tamb (max Tamb +40 ˚C (+104 ˚F) Td)
Operating gases Air, N2, non-toxic gases, non-flammable, inert gases
Inlet/outlet connection 1/4" SWAGELOK
Operating temperature

Ambient temperature: 0 … +40 ˚C (+32 … +104 ˚F)

Gas temperature: max 200 ˚C (392 ˚F) with 2 m (6.6 ft) PTFE tube at 20 ˚C (68 ˚F)

Operating pressure

With pump: 0.6 … 1.2 bara (8.7 … 17.4 psia)

Pump disconnected: 0 … 20 bara (0 … 290 psia)

Battery operation time for pump 8 h continuous
Filter 7 mm (0.28 in) inline filter, 1/4" SWAGELOK SS-4F-7
Wetted parts Stainless steel
Carrying case ABC plastic
Weight 5.5 kg (12 lbs.)
Fuse ratings Glass tube fuse 5 × 20 mm T 2A/250 VAC (Manufacturer: Littlefuse 218002)