Default analog output configuration with Indigo transmitters - DMP5 - DMP6 - DMP7 - DMP8 - DMPX

DMP Series User Guide

Document code
M212357EN
Revision
E
Language
English
Product
DMP5
DMP6
DMP7
DMP8
DMPX
Document type
User guide

When you connect the probe to an Indigo200 or Indigo300 transmitter for the first time, the following analog output values per output parameter are active by default, unless you have ordered the transmitter with preconfigured analog outputs.

Table 1. Default analog output configuration per output parameter
Output parameter Output unit AOUT min AOUT max
Absolute humidity g/m3 0 600
Absolute humidity at NTP g/m3 0 600
Dew/frost point temperature °C −80 100
Dew/frost point temperature at 1 atm °C −80 100
Dew point temperature °C −80 100
Dew point temperature at 1 atm °C −80 100
Dew point temperature difference °C −10 50
Enthalpy kJ/kg −40 1500
Mixing ratio g/kg 0 500
Relative humidity %RH 0 100
Temperature °C −40 60
Water concentration ppmv 0 1 000 000
Water mass fraction ppmw 0 1 000 000
Water vapor pressure hPa 0 1000
Water vapor saturation pressure hPa 0 1000
Diagnostic parameters 1
Sensor saturation ratio % 0 100
Sensor temperature °C −80 220
1 Selectable as output parameters using the Indigo80 handheld indicator and the Modbus protocol.