Wiring - DMP5 - DMP6 - DMP7 - DMP8 - DMPX

DMP Series User Guide

Document code
M212357EN
Revision
E
Language
English
Product
DMP5
DMP6
DMP7
DMP8
DMPX
Document type
User guide
Figure 1. M12 5-pin A-coded male connector pinout
Pin # Function Notes Wire colors in Vaisala cables
1 Power supply

Operating voltage: 15–30 V DC

Current consumption: 10 mA typical, 500 mA max.

Brown
2 RS-485 - White
3 Power GND and RS-485 common Blue
4 RS-485 + Black
5 Not connected Gray
Figure 2. RS-485 wiring

Recommended maximum length of the RS-485 line is 30 m (98 ft).